Skip to content
ACCUMAX Instruments Pvt. Ltd.

Flow Measurement

How Liquid Conductivity Affects Magnetic Flow Measurement

Why electromagnetic flow meters need conductive liquids, what the 20 µS/cm minimum for water means, and how to deal with low-conductivity fluids.

20 April 20266 min readAccumax Engineering Team
How Liquid Conductivity Affects Magnetic Flow Measurement

Electromagnetic flow meters are popular because they are obstruction-free, have no moving parts and measure volume flow regardless of density or viscosity. They do have one firm requirement: the liquid must conduct electricity. Understanding that requirement helps you avoid the most common selection error, which is putting a magnetic meter on a fluid that cannot support it.

The electromagnetic flow meter from Accumax is specified for water with conductivity of at least 20 µS/cm. This article explains what that means in practice.

The principle in brief

According to Faraday's law of induction, a conductor moving through a magnetic field generates a voltage. In the meter, the conductor is the liquid itself. The electrodes detect a very small voltage proportional to average velocity, and the transmitter converts it to flow rate using the known pipe area.

Because the signal is so small, the transmitter's input must be very high impedance. The liquid, the electrodes and the electronics form a circuit, and a poorly conductive liquid weakens that circuit, which leaves the signal exposed to electrical noise.

What conductivity means

Conductivity measures how readily a liquid carries current, expressed in microsiemens per centimetre (µS/cm). It depends on dissolved ions. As a rough orientation, not a specification:

LiquidTypical conductivity behaviour
Ultra-pure or demineralised waterVery low; often unsuitable
Distilled waterLow; usually unsuitable
Typical tap and groundwaterOften hundreds of µS/cm or more
Brackish water, effluentHigher
Dilute acids, alkalis, salt solutionsUsually high
Hydrocarbons, oils, solventsEssentially non-conductive

Actual values vary with source, season and temperature, so measure rather than assume.

The 20 µS/cm rule for water

For Accumax's electromagnetic meter, water needs conductivity of at least 20 µS/cm. Below that, readings may become unstable, jump or drop to zero. Treat the figure as a minimum for reliable operation, and leave margin: if your water measures 22 µS/cm in the lab, seasonal change could push it under the line.

Note that the limit is stated for water. Other liquids can behave differently, so give the supplier your fluid details.

Fluids that cause trouble

  • Demineralised and RO permeate water. Common in boiler feed, pharmaceuticals and electronics. Conductivity is often far too low.
  • Hydrocarbons and solvents. Fuels, oils and many organic liquids are non-conductive; consider other technologies such as turbine, positive displacement or Coriolis.
  • Two-phase flow. Air bubbles or oil droplets in water interrupt conduction and cause noisy readings.
  • Partially filled pipes. The electrodes lose contact with the liquid. A magnetic meter measuring a non-full pipe will read wrongly.

How to check conductivity

  1. Take a representative sample from the actual line, not a bench container left standing.
  2. Measure with a calibrated conductivity meter at known temperature. Conductivity rises with temperature, so note the temperature.
  3. Repeat across different operating conditions, such as after regeneration cycles or seasonal source changes.
  4. Record the lowest value you see and compare with the meter's minimum.

If your site already has analytical instrumentation, a conductivity reading from the process may already be logged. Our pH analyser page covers related online analysis if you are building a wider monitoring picture.

Practical tips for marginal conductivity

  • Choose the shortest, cleanest signal cable and ensure the grounding recommendations are followed. Poor earthing is more damaging at low conductivity.
  • Keep the pipe full. Install in a rising section or a U-bend rather than at the top of a long horizontal run. See flow meter installation and straight run.
  • Ground the liquid using grounding rings or electrodes when the pipe is plastic or lined, so the liquid is at a defined reference potential.
  • Avoid injection points immediately upstream, where chemicals can create layers of different conductivity and cause noisy readings.
  • Consider alternatives when conductivity is truly low. The comparison in electromagnetic vs ultrasonic flow meter is a good start.

Conductivity versus electrode material

Higher conductivity does not mean you can ignore chemistry. Highly conductive fluids are often corrosive, so the liner and electrode selection matters. See choosing liner and electrode materials. Equally, coatings on the electrodes can raise contact resistance, and a meter that once worked on low-conductivity water may stop when deposits build up.

Reading the symptoms

SymptomPossible conductivity-related cause
Reading jumps with no process changeConductivity close to the minimum, noise pickup
Reading falls to zero at low flowEmpty-pipe detection or lost electrode contact
Works in summer, fails in winterTemperature-dependent conductivity
Reading drifts over weeksElectrode coating

Before blaming the meter, always verify the liquid. For wider diagnosis, see diagnosing unstable flow readings.

Acceptance checks before handover

Run a documented comparison under representative conditions and retain the original observations. Check that the local display, transmitted value and stored record use the same units and scaling. Include an approved interruption test, then verify recovery without silently resetting totals or losing the configuration. Assign responsibility for reviewing alarms and unexplained gaps. These checks establish a baseline for maintenance rather than proving that the installation will remain correct indefinitely.

Frequently asked questions

What is the minimum conductivity for the Accumax electromagnetic flow meter?

For water, conductivity must be at least 20 µS/cm. Allow a safety margin for variation in your process.

Can an electromagnetic meter measure oil or fuel?

Generally not, because these liquids are essentially non-conductive. Another measurement principle should be selected.

Does temperature change conductivity?

Yes. Conductivity usually increases with temperature, so measure at representative conditions, including the coldest operating case.

Conclusion

Conductivity is the gatekeeper for magnetic flow measurement. Verify it with a real sample, keep a margin above the 20 µS/cm minimum for water, keep the pipe full, and ground properly. If the liquid does not qualify, choose a different principle rather than forcing the fit.

Unsure whether your liquid is suitable? Send us the details through our contact page.

Written by the Accumax Engineering Team

The people who design, build and calibrate Accumax instruments in Gandhinagar since 1992.

FAQs

Frequently asked questions

Can’t find your answer? Our engineers reply on WhatsApp at +91 88496 70588.

What does this article explain about Liquid Conductivity Magnetic Flow Measurement?

This article discusses liquid conductivity magnetic flow measurement and its practical instrumentation considerations. For a specific instrument selection or unlisted value, For configuration or values not stated here, contact Accumax engineers on WhatsApp +91 88496 70588 or at sales@accumaxinstruments.com.

What should an engineer consider when applying this guidance?

The article addresses liquid conductivity magnetic flow measurement in an instrumentation context. Check the actual process conditions and current requirements before finalising a design; For configuration or values not stated here, contact Accumax engineers on WhatsApp +91 88496 70588 or at sales@accumaxinstruments.com.

Does this article give a single specification for every installation?

No universal site-specific value should be inferred from general guidance. Confirm the required configuration with Accumax engineers on WhatsApp +91 88496 70588 or sales@accumaxinstruments.com.

How can I confirm the product details related to Liquid Conductivity Magnetic Flow Measurement?

Use the article as technical guidance and verify the relevant product data for your application. Accumax engineers can review details that are not stated in the content; For configuration or values not stated here, contact Accumax engineers on WhatsApp +91 88496 70588 or at sales@accumaxinstruments.com.

Where can I ask about the recommendations in this article?

Contact Accumax Instruments Pvt Ltd, Gandhinagar, Gujarat, for application-specific guidance. Call or message WhatsApp +91 88496 70588 or email sales@accumaxinstruments.com.